DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

DR0566 - Soft and Antenna Cover 21 cm Digital Detector
1 Personnel Requirements
Required Persons 1 Preliminary Reqs 5 mins Procedure 30 mins Finalization 5 mins

2 Preliminary Requirements
2.1 Tools and Test Equipment
Item Standard toolkit with metric wrenches Qty 1 Effectivity Part# Manuf -

2.2 Consumables
No consumables required.

2.3 Replacement Parts
Item Antenna Cover Soft Cover Front 21 cm Lift Bottom Cover with Bumper Qty 1 1 1 Effectivity Part# 5134571 2360959-2 5214961 Manuf -

2.4 Safety

RISK OF ELECTRIC SHOCK HAZARDOUS VOLTAGE PRESENT

BEFORE ANY INTERVENTION SWITCH OFF THE SYSTEM (REFER TO PROTECTION AGAINST ELECTRICAL HAZARDS IN SAFETY AND REGULATORY FOLDER OF YOUR INTERACTIVE SERVICE INFORMATION (ISI))

2.5 Required Conditions

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 1 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

Condition Remove the elevator cover Remove power from the system

Reference Refer to DR0565 - Elevator Cover - 21 cm Digital Detector REG0049 - Lockout Tagout Procedures

Effectivity -

3 Procedure
3.1 Disassembly
1. Illustration 1: Remove the anti-scatter grid .

2. Illustration 2:

Remove the C-arc end cover, by dismantling the two screws

.

3.

Disconnect the air flow tube

between cover and lift.

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 2 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

Illustration 3:

4. Illustration 4:

Unscrew the five screws

securing the elevator cover.

5.

Remove the elevator cover

.

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 3 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

Illustration 5:

6. Illustration 6:

Disconnect the five plugs

from the capacitive sensor and the air flow tubes

.

7.

Disconnect the switch ring cover

.

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 4 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

Illustration 7:

8. Illustration 8:

Unscrew the three screws

securing the bottom cover with bumper.

9.

Unscrew the four screws

securing the switch ring.

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 5 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

Illustration 9:

10. Illustration 10:

Loose switch ring bracket

secured by two screws

and disengage the switch ring

.

Last Screws to remove the Switch Ring/Soft Cover/Antenna Cover Assembly.

11.

Unscrew the 3 antenna cover screws

.

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 6 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

Illustration 11:

12. Illustration 12:

Remove the switch ring, soft cover and antenna cover assembly

.

13.

Unscrew the four screws

and remove the Soft and Antenna covers.

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 7 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

Illustration 13:

14.

Separate the Soft

and Antenna covers

(refer to Illustration 13 ).

3.2 Reassembly
1. Reassemble in the reverse order of Disassembly procedure.

4 Finalization
1. 2. 3. 4. 5. Re-install the elevator cover. Refer to DR0565 - Elevator Cover - 21 cm Digital Detector . Remove Gantry rear trim cover. Refer to DR0012 - L-Arm Front and Rear Trim Covers . Power on the system. Refer to REG0049 - Lockout Tagout Procedures . Check level 1 and level 2 collision detection at system level by pressing the bumper. On frontal vacuum board, remove J4 jumper and insert it on J6 to launch autotest.

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 8 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

Illustration 14:

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 9 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

Illustration 15:

6. 7. ? ? NOTE:

When pump starts, remove J6 jumper. At the end of the test (this check takes approximately 3 minutes) check Leds status: DS4 and DS5 must be ON. DS3 and DS6 must be OFF.

If DS3 blinks, there is a leakage in pneumatic circuit. Check the complet pneumatic circuit (Pressure Control Module, bumper, hoses and hoses T-connectors). 8. When test ok, re-install jumper between 2 and 3 of J4.

REFER TO REG0049 - LOCKOUT TAGOUT PROCEDURES
9. 10. Remove power from the system. Re-install Gantry rear trim cover. Refer to DR0012 - L-Arm Front and Rear Trim Covers .

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 10 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Direction 2360772-11-200 , Revision 1

11.

Power on the system. Refer to REG0049 - Lockout Tagout Procedures .

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.

Page 11 of 11

DR0566 - Soft and Antenna Cover 21 cm Digital Detector GE Medical Systems Blank page. Direction 2360772-11-200 , Revision 1

Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Renewal Parts and Disassembly / Reassembly Procedures Copyright ? 2007 by General Electric Company Inc. All Rights Reserved.